Clinical Biochemistry Lab Outline

Introduction

The Role of the Lab in Monitoring Health and Disease

When a person goes to their doctor with a concern about their health, it is relatively common to leave the doctor’s office with a laboratory requisition in-hand. When a person goes to the emergency room, one of the first medical procedures performed is often bloodwork. Physicians may use the results of a laboratory test as a starting point from which to further make decisions about next steps in a person’s healthcare – what tests to order next, what specialists to send them to, and/or what medications they should prescribe. Laboratory tests are the most frequently ordered diagnostic procedure, and the clinical laboratory remains an essential tool in the practice of diagnostic medicine (Ngo et al., 2017; Hilbert et al., 2022). The purpose of the clinical laboratory is to provide health care professionals and patients with the necessary information to:

  • detect a disease or the predisposition to a disease;
  • confirm or reject a diagnosis;
  • establish prognosis;
  • guide patient management; and
  • monitor the efficacy of therapy.

They may also use clinical testing as a baseline to confirm or monitor a person’s good health status as part of routine check-ups (Hilbert et al., 2022).

The most common biological samples tested in clinical biochemistry labs are blood and urine. There are a wide array of tests that can be conducted on even just these two types of samples. The vast majority of biochemical analyses are quantitative, meaning that the results are expressed in numbers, amounts or quantities. In contrast, some results can be more qualitative which provide a description of characteristics (e.g. appearance, colour, smell, etc.), though these observations are usually in complement to the quantitative results. For example, a basic examination of urine would include a description of its appearance, colour, and odor, while also being tested for quantitative characteristics like pH, specific gravity, and quantification of the components requisitioned by the physician (Riley & McPherson, 2022). The measurement of these analytes would be compared to a reference interval, which is the representative “normal” range for 95% of the general population.

For the purpose of this lab, you will be working as a laboratory technician who is conducting urinalysis testing on a sample submitted from a family doctor’s office. A urinalysis evaluation may include any or all of the following:

Table 2: Summary of urinalysis evaluation.

Evaluation Criteria
Gross/physical examination Appearance, odor, volume, specific gravity and osmolality
Chemical screening pH, protein, glucose (and other sugars), blood, bilirubin, urobilinogen, indirect infection testing (nitrites, leukocytes, PCR) other miscellaneous when indicated
Infectious disease screening Bacterial culture, antibiotic susceptibility testing, PCR or DNA sequencing
Microscopic examination Atypical inclusions, e.g. blood cells, fat bodies, casts, crystals, tumor cells, bacteria, parasites

 

The primary care physician (i.e family doctor) has requested a chemical urinalysis test. This involves the use of carefully calibrated urine test strips. The results can be interpreted semi-quantitatively through visual inspection and comparison to a chart containing results ranges, or quantitatively using a urine analyzer that can quantify the colorimetric results from the different assay pads on the test strip. Urine analyzers can process single strips at a time, or can be automated to process multiple samples at a time. We don’t have a urine analyzer in our facility, so you will quantify the analyte of interest in the urine sample using the semi-quantitative approach of  visually comparing the results on the test strip to visual references and quantitatively using a biochemical assay. Quantification will require the creation of a standard curve and the use of a plate reader spectrophotometer. The type of analytical instrument to be used and what will be quantified are dictated by the analyte type, what biochemical assay(s) you are using, and what is the resultant measurable output from the assay.

The purpose of a standard curve is to define a measured signal of interest against a set of reference samples with known concentrations. They can then be used to determine the unknown concentration of an analyte using the equation from the standard curve.

Linear graph comparing concentration and absorbance. Representative of a standard curve.
Figure 1: An example of a standard curve for spectrophotometry. A set of standard samples with known concentration are measured using a spectrophotometer. The results are graphed to compare the absorbance at a specific wavelength with their known concentrations. Unknown concentrations of the same analyte can be determined by measuring their absorbance. Note that standard curves always have units indicated on the measures for the X and Y axis, which will be dictated by what wavelength you are measuring your sample at and what the concentration is of your standard samples.

 

The relationship between concentration and absorbance is dictated by the Beer-Lambert law, which states that concentration, path length (i.e. the distance light has to travel within a sample) and molar absorptivity are all directly proportional to the absorbance, as represented by the equation:

A = εbc,

where A = absorbance, ε = molar absorptivity (how strongly the substance absorbs light at a given wavelength), b = distance of light traveled, c = concentration of the solution

This can be re-interpreted as the equation of a straight line (y = mx +b) with a y-intercept of zero, i.e. y = mx, where you would solve for x, in this case the concentration of a measured unknown. Here is a video that nicely explains how absorbance data is interpreted from a standard curve:

Here is the information you have been provided about the patient that you may find relevant to your analysis:

Ordering Physician: Dr. Nematode

Sex: Male

Date of Birth (DOB): 26/04/1962

Age: 64

Collection Date: *the same day you received the sample*

Collection Time: 8:30 AM

Requisition: Urinalysis (chemical); must include glucose quantification

Physician Notes: Suspected type 2 diabetes; testing for glycosuria because blood sample could not be collected to test for hyperglycemia.

Table 2 provides reference ranges for glycosuria that you should compare your urine sample to. In general, glucose should not be detectable in urine. Levels above 100 mg/dL are considered abnormal. Glycosuria is the presence of excessive sugar (glucose) in the urine, typically exceeding 0.25 mg/mL (25 mg/dL), occurring when blood sugar levels are too high or when kidneys cannot reabsorb glucose properly. Sugar does not normally pass into the urine until blood glucose levels are 180 mg/dL or higher (Liman & Jialal, 2023; Hieshima et al. 2020).

 

Table 2: Reference ranges for glucose levels in urine (Mundt & Shanahan, 2016).

Grade Concentration (mg/dL) Concentration (mmol/L) Clinical Meaning
Negative (normal) 1 - 15 0 - 0.8 Normal finding; no significant glucose in urine.
Trace (atypical) ~100 ~5 Borderline or low-level presence; repeat or check blood sugar.
1+ (mild) ~250 ~15 Mild glycosuria; suggests elevated blood sugar or reduced renal threshold.
2+ (moderate) ~500 ~30 Moderate glycosuria; commonly linked to hyperglycemia.
3+ (high) ~1,000 ~60 Marked glycosuria; strong indicator of high blood glucose.
4+ (very high) ≥ 2000 ≥ 110 Severe glycosuria; heavy glucose spillage into urine.

Lab Overview and Timeline

In this lab, you will investigate the concentration of glucose in a urine sample using a urine test strip (also known as a dipstick) as well as measure the concentration using a plate reader spectrophotometer. Part of this exercise will involve creating a standard curve using known concentrations of glucose.

This lab will be performed in PAIRS, however there will be some parts where you will do parts of the procedure with another pair of students at your station.

Time (min) What will you be doing? Total Time (min)
0 - 20 Overview of lab activities by the TA.
Fill in pre-lab of lab notebook.
15
20 - 35 Lab Part 1: Semi-Quantitative Urinalysis Using a Test Strip
Lab Protocol 1: Performing Urinalysis using a Test Strip
15
35-120 Lab Part 2: Dilution of Standards and Test Sample
Lab Protocol 2a: Creating Glucose Standards
Lab Protocol 2b: Diluting your Unknown
Lab Part 3: Quantitative Urinalysis using a Colorimetric Glucose Quantitation Assay
Lab Protocol 3: Preparing Samples and Performing Glucose Quantitation Assay
90
120-170 Finish lab notebook and submit on Crowdmark. 50

 

Safety Reminders

You should always exercise caution and use common sense when working in any lab. In the lab this week, these are a few of the activities where extra care is warranted:

  • Personal devices
    • You will be allowed to use your cellular phone to take photographs of your urinalysis test strips for your ELN and DLR. Remember to remove your gloves before handling your personal device, and put the device away once you are finished with Protocol 1 so as not to contaminate your phone with lab reagents.
  • Pipetting reagents
    • You will be mixing reagents in small volumes using a micropipette. Micropipette tips are considered as sharps and should be disposed of in the provided sharps container.
    • You must wear gloves, a lab coat, and chemical splash goggles when working with the reagents provided. The GOD-POD reagent has a small amount of dimethyl sulfoxide (DMSO) which can cross biological membranes. You may also notice a sulfur smell. Use caution when working with this reagent.

Lab Part 1: Semi-Quantitative Urinalysis Using a Test Strip

Purpose

To evaluate a urine sample across seven parameters (glucose, protein, ketones, pH, blood, nitrites, and leukocytes) and determine the amount of glucose using a semi-quantitative method.

Summary of Activities

You will be using the BTNX Rapid Response 7-parameter urinalysis reagent strip to do an initial assessment of your urine test sample.

Before you begin, ensure you are wearing the appropriate personal protective equipment (PPE): gloves, lab coat and chemical splash goggles.

Materials:

  • Unknown urine sample (approximately 10 mL)
  • Urinalysis test strip
  • Phone for taking images

Lab Protocol 1: Performing Urinalysis using a Test Strip

  1. Document which unknown urine sample you have at your station (e.g., Unknown #1)
  2. Mix your unknown urine sample for 5 seconds using the vortex.
  3. Label a microcentrifuge tube “unknown.” Remove 500 uL of your urine sample and pipet it into this tube. Set this off to the side for Part 2.
  4. Watch the video below on how to properly use the Rapid Response urinalysis test strips.

  1. Dip the test strip provided into the remaining unknown urine sample provided to you. Carefully wipe the edge of the test strip as seen in the video and place it on a paper towel. Be sure that the reagents from the pads on the strip do mix into each other.
  2. Start the timer and wait 30 seconds. Document your result by taking a photograph of the test strip with one of your cell phones. You will include this photograph in your ELN.
  3. Read the concentration range for glucose by comparing the glucose results of your test strip to the chart on the test strip bottle or the colour chart that is provided.
  4. Take another photograph of the test strip after 60 seconds and read the concentration ranges for protein, pH, blood, and nitrite. Be sure to write down the concentration ranges for each of the results.
  5. Take one last photograph of the test strip after 120 seconds and read the concentration range for leukocytes. Be sure to write down the concentration range for this result.
  6. Add the three photographs of your test strip to your ELN. Make sure to be specific in your ELN about the glucose concentration identified by the test. Save all three images for use in your DLR.
  7. Dispose of your test strip in the biohazard waste container.

Lab Part 2: Dilution of Standards and Test Sample

Purpose

To generate a standard curve and quantify the amount of glucose in your unknown urine sample using a colorimetric quantitation assay.

Summary of Activities

Students will perform serial dilutions of a known quantity of glucose and use a plate reader spectrophotometer to measure the absorbance of the standard samples as well as the unknown urine sample. These measurements will then be used to generate a standard curve. The standard curve will then be used to calculate the concentration of glucose in the unknown sample.

Once you have obtained a glucose concentration range from your urinalysis test strip, you will conduct a quantitative assay to confirm the glucose concentration in the urine sample. To do this, you will need to perform a standard curve with known concentrations of glucose. You will then analyze your unknown and use linear regression analysis to identify the concentration of glucose in the urine sample.

In biological fluids such as blood and urine, glucose can be measured and detected in multiple ways. Commonly, quantification is coupled with enzymes that break down the molecule of interest (in our case glucose) and produce a secondary product that can then be further coupled with a colorimetric or fluorescent output that can be quantified using a spectrophotometer. Examples of glucose quantification assays include the hexokinase assay, the glucose-oxidase-peroxidase assay, and the glucose dehydrogenase (GDH) assay.

The method we are using for this lab are based on a widely used enzyme-coupled colorimetric assay called the glucose oxidase-peroxidase (GOD-POD) assay. It uses glucose oxidase-based enzyme coupled reactions to detect glucose through the production of hydrogen peroxide, which is monitored by a peroxidase substrate, producing a red coloured product. This substrate can then be read by an absorbance microplate reader at 570 nm.

 

Reaction equations for the glucose oxidase-horseradish peroxidase reaction.
Figure 2: The glucose oxidase (GOD) - horseradish peroxidase (POD) reaction.
For this assay the final oxidized Amplite(R) reagent produces a red dye whose absorbance can be measured at 570 nm.

 

The reagents we are using are very sensitive and detect glucose in the micromolar (μM) concentration range. The urine sample you have tested in Part 1 will have tested positive in the millimolar (mM) range, meaning that we will need to dilute the urine sample 1000x before we can test it using the GOD-POD reagent.

You should review Appendix C: Preparing Chemical Solutions, specifically the “Diluting and Dilution Factors” information box (in blue) prior to performing the following task.

Materials:

  • Unknown urine sample
  • Deionized (DI) water
  • Assay buffer
  • GOD-POD reagent
  • Microcentrifuge tubes
  • Vortex
  • P1000, P200 and P20 micropipettes
  • 96-well plate
  • Timer

Lab Protocol 2a: Creating Glucose Standards

You will be preparing a standard curve with 5 data points: one blank (no glucose) and 4 different concentrations: 12.5 μM, 25 μM, 50 μM, 100 μM. You will be performing 1:1 (i.e. 2x) serial dilutions to achieve these concentrations.

  1. Prepare 5 microcentrifuge tubes and label them as follows: 0, 12.5 μM, 25 μM, 50 μM, 100 μM.
  2. Pipet 200 μL of DI water into the “0” tube and set it aside. This will be your blank.
  3. Arrange the rest of your tubes in a row from highest to lowest concentration. Add 200 μL of DI water to each tube.
  4. Refer to Figure 3 for visual guidance on performing the serial dilutions. The steps are outlined as follows:
    • Gently mix the 200 μM stock solution for 5 seconds using the vortex. Transfer 200 uL from this tube to your 100 μM tube. Mix briefly.
    • Transfer 200 μL from the 100 μM tube to the 50 μM tube. Mix briefly.
    • Transfer 200 μL from the 50 μM tube to the 25 μM tube. Mix briefly.
    • Transfer 200 μL from the 25 μM tube to the 12.5 μM tube. Mix briefly.
Serial dilution of glucose stock solution. The image shows 5 microcentrifuge tubes with arrows showing how to transfer solutions between the tubes to perform serial dilution.
Figure 3: Dilution procedure for making stock glucose concentrations using serial dilution.

 

If completed correctly, you should have 200 μL in the 100 μM, 50 μM, and 25 μM tubes, and 400 μL in the 12.5 μM tube. You have now completed your 1:1 dilution series.

Lab Protocol 2b: Diluting your Unknown

The unknown urine sample will have a glucose concentration in the mM range. It needs to be diluted down to the μM range (i.e., a 1000x dilution). This can be achieved in one step or in multiple steps by serial dilution. We will be doing this in two steps, first with a 100x dilution, then with an additional 10x dilution (see Figure 4).

Serial dilution of unknown urine sample. The image shows a test tube and 3 microcentrifuge tubes, with arrows showing how to transfer solutions between the tubes to perform serial dilution.

Figure 4: Dilution procedure for your unknown urine sample using a 100x and 10x serial dilution. The final solution (Dil 2) will have been diluted 1000x from the original sample.

  1. Prepare 2 microcentrifuge tubes and label them as follows: “Dil 1”, “Dil 2”
  2. Retrieve the microcentrifuge of unknown urine prepared in Part 1. Mix briefly.
  3. Prepare Dilution 1 (Dil 1) which will dilute your urine sample 100x:
    • Pipet 990 μL of DI water into the tube labeled Dil 1.
    • Pipet 10 μL of your unknown urine sample into the tube labeled Dil 1. Mix briefly.
  4. Prepare Dilution 2 (Dil 2) which will dilute your urine sample a further 10x, achieving a 1000x dilution overall:
    • Pipet 180 μL of DI water into the tube labeled Dil 2.
    • Pipet 20 μL from Dil 1 into the tube labeled Dil 2. Mix briefly.
  5. You will use the sample in Dil 2 to perform your glucose quantification in Part 3.

 

Lab Part 3: Quantitative Urinalysis using a Colourimetric Glucose Quantitation Assay

For this part, you will be mixing your glucose standards and diluted unknown urine sample with the glucose oxidase-peroxidase (GOD-POD) reagent and quantify the glucose in the samples using a plate reader, which is a spectrophotometer that reads samples from a 96-well plate.  This will involve transferring your samples in triplicate to the 96-well plate and incubating them for a period of time in the dark at 37ºC. Your TA will assist you with using the plate reader to obtain your results.

Each pod will share a 96-well plate. Each pair of students will have a total of 18 samples to load on the plate, where you will have:

  • 3 wells of 0 glucose (your blank)
  • 3 wells each of 12.5 μM, 25 μM, 50 μM, and 100 μM glucose standard
  • 3 wells of your diluted unknown glucose sample (Dil 2).

These are called technical replicates. Their purpose is to evaluate the reproducibility of the assay or technique. This is different from a biological replicate, which is when distinct biological samples are evaluated in parallel to determine whether a biological phenomenon or experimental effect is generalizable to different organisms or sets of biological samples.

NOTE: The GOD-POD reagent is light-sensitive, therefore you want to work efficiently once you have begun loading your 96-well plate to minimize light exposure as much as possible.

This procedure will be performed in PAIRS, however 2 pairs of students will share one 96-well plate.

Lab Protocol 3: Preparing Samples and Performing Glucose Quantitation Assay

  1. Using the sample 96-well map provided at your bench, plan out where you intend to load each of your samples. It is typical to load each replicate side-by-side. Take a photo of your map to include in your ELN. Alternatively, you can use this online editable plate map to generate a PDF of your layout.
  2. Make your GOD-POD reagent. It must be made fresh just before you set up your experiment. Be sure to change tips between each reagent. Into the tube labeled “GOD-POD” add the following:
    • 956 μL assay buffer
    • 20 μL HRP solution
    • 20 μL GO solution
    • 4 μL Amplite® Red solution – you will obtain this from your TA.
  3. Carefully and quickly pipet 50 μL of GOD-POD reagent into each of your mapped out wells. You may use the same micropipette tip for all 18 wells.
  4. Pipet 50 μL of the 0 glucose sample into its designated well. Mix by pipetting up-and-down gently 5 times. Change the micropipette tip and repeat for the other 2 wells.
  5. Repeat step 4 for each of the glucose standards (i.e., 50 μL of standard into each of their 3 designated wells)
  6. Repeat step 4 for your Dil 2 unknown glucose sample (into each of the 3 designated unknow wells).
  7. If you are the first pair from your pod to load the plate, take a piece of aluminum foil and carefully cover your wells give the plate to the other pair of students in your pod so they can load their side of the plate.
  8. Once all samples have been loaded onto the 96-well plate, cover your plate and put it in the 37°C incubator for 20 minutes.
  9. When your incubation is finished, bring your plate to the TA. They will load your plate into the plate reader and share your results in the form of an Excel spreadsheet.
  10. Upload a PDF copy of your plate reader results into your ELN.
  11. Disposal procedures are as follows. Place your reagent tubes in the micropipette rack on the bench in the middle of the room (the one closest to the cubbies). Place your 96 well plate on the same bench in the middle of the room on the bench top protector pad (the soft white pad with the blue border). Empty your unknown urine tubes in the waste bottle that is found on the middle bench. Ensure the bottle is on the bench top protector pad. Use the funnel if needed. Empty the ice from your reagents cup in the sink. Wash your hands before you leave the lab room.

Lab Exit Checklist

  • Completed urinalysis using the test strip, took a picture of your results for your ELN and DLR, and disposed of the used test strip in the biohazardous waste
  • Performed serial dilutions of glucose standards and unknown urine sample
  • Mapped a 96-well plate for where to load samples.
  • Obtained results from the plate reader spectrophotometer. Saved your results for submission as part of your ELN, and for later used during completion of your DLR.
  • Submit your ELN on Crowdmark before leaving the lab.
  • Ensured that all images and spectrophotometer data for the DLR assignment has been saved to your OneDrive(s).

References

ChemistNate. Beer’s Law: Calculating Concentration from Absorbance. [Video]. YouTube. https://youtu.be/A5RtV1tB-Qo?si=2bBl5h5GZWO7sAAs

Hieshima, K., Sugiyama, S., Yoshida, A., Kurinami, N., Suzuki, T., Ijima, H., Miyamoto, F., Kajiwara, K., Jinnouchi, K., Jinnouchi, T., & Jinnouchi, H. (2020). Elevation of the renal threshold for glucose is associated with insulin resistance and higher glycated hemoglobin levels. Journal of diabetes investigation, 11(3), 617–625. https://doi.org/10.1111/jdi.13191

Hilbert, T., Kurec, A. & Lifshitz, M.S. (2022). General concepts and administrative issues. In R.A. McPherson & M. Pincus  (Eds.). Henry’s Clinical Diagnosis and Management by Laboratory Methods (24th ed). Elsevier. DOI: 10.1016/B978-0-323-67320-4.01002-6.

Liman, M.N.P. & Jialal, I. (2023). Physiology, Glycosuria. StatPearls. https://www.ncbi.nlm.nih.gov/books/NBK557441/

Mundt, A. L. & Shanahan, K. (2016). Graff's Textbook of Urinalysis and Body Fluids (3rd ed., pp 93-94). Jones & Bartlett Learning.

Ngo A., Gandhi P. & Miller G. (2017). Frequency that laboratory tests influence medical decisions. Journal of Applied Laboratory Medicine, 1(4), 410-414. DOI: 10.1373/jalm.2016.021634

Pape, P.T., Sharp, V.J.A., Rockafellow, J. (2020). Urine Dipstick: An Approach to Glucosuria, Ketonuria, pH, Specific Gravity, Bilirubin and Urobilinogen – Undeniable Chemistry. In: Sharp, V., Antes, L., Sanders, M., Lockwood, G. (Eds). Urine Tests. Springer, Cham. https://doi.org/10.1007/978-3-030-29138-9_7

Rapid Response. How to use Rapid Response urinalysis test strips. [Video]. YouTube. https://www.youtube.com/watch?v=Hz3piYV9v_M&t=13s

Riley R.S., McPherson R.A. (2022). Basic examination of urine. In: R.A. McPherson & M.R. Pincus (Eds.) Henry's Clinical Diagnosis and Management by Laboratory Methods (24th ed). Elsevier. DOI: 10.1016/B978-0-323-67320-4.01002-6.

Urinalysis Reagent Strips, 7 Parameters. (n.d.). BTNX. Retrieved September 28, 2026 from https://www.btnx.com/product/urinalysis-reagent-strips-7-parameters

 

definition

License

Icon for the Creative Commons Attribution 4.0 International License

LIFESCI 2L03: Living Systems Laboratory Course Pack (8th Edition) Copyright © by Ryan Belowitz; Ana Tomljenovic-Berube; and Devon Jones is licensed under a Creative Commons Attribution 4.0 International License, except where otherwise noted.